[FACA-261] Fully Automatic Chemistry Analyzer 250 Tests per Hour - Labomed Inc.

[FACA-261] Fully Automatic Chemistry Analyzer 250 Tests per Hour

 

This automatic biochemistry analyzer comes with varying testing capabilities that conducts 250 tests per hour with true random access and 10 filters: 340nm, 405nm, 450nm, 546nm, 578nm, 620nm, 700nm, 800nm. Additional filters can be added.

Features

  • Each set of FACA will come with one video to train in the installation, operation, and maintenance of the biochemistry analyzer.
  • Random-Access with open system.
  • 80 Cuvette positions, semi disposable plastic for min 30,000 tests/80 Cuvettes.
  • Cuvette Washing Station - 8 channels automatic washing station, 1 for detergent wash (connect deteregnt bottle), 2-6 channels wash by water, �7 channel for first empty, 8 channel for final empty
  • Reliable cooling system for reagent chamber - Peltier pad
  • Independent mixer
  • On-board refrigeration of reagents with independent switch [9�C below ambient temp.]
  • Automatic pre- and post-dilution.
  • External sample barcode reader, with positive identification of samples.
  • Primary tubes up to 16 x 100 mm and sample cups.
  • Preinstalled & validity limits for methods and reactions programmable.
  • Intelligent flagging system and extensive error log.
  • Reagent integrity check & extensive QC monitor.
  • Water consumption: 3 L/hr
  • Complete reaction curve & statistics for every sample/test-not only QC.
  • Dry incubation temperature control: �1.0�
  • LIS software optional - must be purchased separately
  • Reads serum and urine samples
  • Absorbance alert provided when the Absorbance is abnormal
  • ISE Module available upon request.
  • All FACA Fully Automatic Biochemistry Analyzers have 2 different sample plates:
    1. Standard sample plate for use with serum cup (serum order)
    2. Test tube sample plate for use with 13x75 - 13x100mm test tubes (that is optional)

AnalyzerType:

  • Fully Automated, Bench Top/Floor Top, Discrete, Random-Access with STAT functionality open system Clinical Chemistry Analyzer.

Throughput:

  • 250 tests / hour (without ISE)

On-board Samples:

  • At least 60 sample positions for Primary tubes & Sample cups and option for additional (20 20) =40 positions for Primary tubes & Sample cups.
  • Sampling volume:�2 to 300 �l
  • Automatic Pre-dilution & Post-dilution of samples.
  • Max 30 different parameters can be performed simultaneously.

On-board Reagents:

  • Min 40 positions�with on board refrigeration�at 9 �C below ambient tempwith independent switch for 24 hours reagent cooling even if the machine fails to run.
  • Reagent (R1-R2) Volume: 5 to 350�l

Optical Unit:

  • Wavelength Spectral Range:340-900 nm�(extendable)
  • 8 positions filter wheel with 7 pre-installed supplied filters & 01 free position.
  • 8 filter with 340-900nm
  • Absorption Range (Linearity):0.000 to 2.500 Abs
  • Accuracy:��� 2 nm (wavelength)
  • Band Pass Precision:��5 mm

Reaction Unit:

  • Reaction Volume:210-350 �l
  • Reaction wells positions: 80 Cuvette positions, semi disposable [> 30,000 tests].
  • Incubation temperature & Time:�37�C � 0.2�C & 792 sec.�max. incubation�including reading time.

Wash station:

  • 8-step wash station with 8 dispensing needles
  • Two types of Systemic (20L) & special washing solution (2L) and 20L waste with level sensor.
  • 2nd cleaning solution & extra washing definable in method.
  • Water consumption: max� 3 L/hr

Liquid Handling:�����������������������������������������������������

  • Liquid transportation through pipetting arm with�capacitive liquid level sensors.
  • Needle shock detector &�vertical collusion detection for probe.

Methodology:

  • Measuring modes:�Routine chemistry, Turbidimetry.
  • Analysis modes:�End Point (bichromatic) & Differential end point with self blank, Fixed Time & Kinetic (Bichromatic).
  • Calculation modes:�Factor, Multi Standard, Single Standard.
  • Calculation algorithms:�Linear (factor, linear, linear regression, average) & Non linear (cubic spline, poly-linear, logit-log).
  • Programmable Test:�50 active Programmable Test methods on instrument with unlimited programmable profiles.
  • Standard Calibration Method:�Up to 8 calibrator standards per test.

Data Processing & Quality Control (QC):

  • Memory capacity:��Storage in HDD
  • Memory:�For Sample results, calibration, patient data, QC data, error-log, absorbance curves.
  • Reports:�For Patient, single test, complete sample, work sheet, method and QC�s, calibration curves, kinetics, continuous printing.
  • QC Level:�Up to max 3 levels per test
  • QC Module:�Levey-Jennings plots, Westgard multirules, SD, CV% & Mean.
  • Warnings:�Analytical limits, reagent and reaction integrity check (blank, linearity, substrate depletion, reaction OD etc) with extensive QC monitor.

Unique versatile user software:

  • New versatiletouch screen user software with intuitive smart graphical user interface.
  • Extended walk-away capability with free choice of execution order.
  • ASTM Bi-directional LIS,�Ethernet of external PC, polling mode on LAN.

Reagents: